{"id":4599,"date":"2026-08-17T12:57:27","date_gmt":"2026-08-17T12:57:27","guid":{"rendered":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india"},"modified":"2026-11-17T03:41:21","modified_gmt":"2026-11-17T03:41:21","slug":"body-composition-analyzer-for-hospitals-india","status":"publish","type":"post","link":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india","title":{"rendered":"Body Composition Analyzer for Hospitals: Clinical Spec Guide"},"content":{"rendered":"<h2>Specifying a Body Composition Analyzer for a Hospital Is a Different Problem<\/h2>\n<p>A gym needs a fast test and a member-friendly printout. A hospital needs something else entirely: outputs that hold up in a clinical decision, fluid-status data that changes management, and a device that produces the same value on the same patient on Tuesday as it did on Monday.<\/p>\n<p>Buying a <strong>body composition analyzer for a hospital or clinic<\/strong> on gym specifications is the most common and most expensive procurement error in this category. This guide sets out what clinical settings actually require, which departments use which outputs, and what to put in the tender.<\/p>\n<figure class=\"ib-auto-img\" style=\"margin:28px auto;text-align:center;max-width:640px;\"><img decoding=\"async\" src=\"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/inbody-770-clinical-analyzer.jpg\" alt=\"InBody 770 clinical body composition analyzer used in Indian hospitals\" width=\"640\" loading=\"lazy\" style=\"max-width:100%;height:auto;border-radius:10px;\" \/><figcaption style=\"font-size:13px;color:#666;margin-top:8px;\">InBody 770 clinical body composition analyzer used in Indian hospitals<\/figcaption><\/figure>\n<h2>The Clinical Outputs That Justify the Specification<\/h2>\n<p>Three outputs separate a clinical-grade analyser from a professional fitness one. Each maps to a specific clinical use.<\/p>\n<h3>Phase Angle<\/h3>\n<p>Phase Angle is derived from the ratio of reactance to resistance in the impedance measurement. It reflects cell membrane integrity and cellular health. Low values indicate cellular stress \u2014 seen in malnutrition, cancer cachexia and chronic disease \u2014 while higher values indicate robust cellular health.<\/p>\n<p>It is used clinically in oncology, nephrology and critical care to assess nutritional status and as a prognostic indicator. Only higher-end models measure it; the InBody 770S reports both whole-body and segmental Phase Angle. Our detailed explainer is at <a href=\"\/blog\/phase-angle-explained-what-is-a-good-phase-angle-and-why-it-matters\">phase angle explained<\/a>.<\/p>\n<link rel=\"stylesheet\" href=\"https:\/\/www.inbody.in\/blog\/wp-content\/mu-plugins\/ib-figures\/figures.css?ver=1794886854\"><figure class=\"ibf ibf-illus\" data-ibf><div class=\"ibf-art ibf-a-up\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 640 400\" stroke-linecap=\"round\" stroke-linejoin=\"round\" role=\"img\" focusable=\"false\"><title>Phase angle reflects cell membrane health: low in malnutrition and cachexia, higher in healthy cells<\/title>\n<circle cx=\"440\" cy=\"210\" r=\"150\" fill=\"#f6e4e9\"\/>\n<path d=\"M90 250c26 76 214 76 240 0\" fill=\"none\" stroke=\"#6b6264\" stroke-width=\"3\" stroke-dasharray=\"8 8\"\/><path d=\"M318 256l12-8 4 14\" fill=\"none\" stroke=\"#6b6264\" stroke-width=\"3\"\/>\n<g stroke=\"#1c1718\" stroke-width=\"3\"><circle cx=\"210\" cy=\"190\" r=\"92\" fill=\"#f6e4e9\"\/><circle cx=\"210\" cy=\"190\" r=\"74\" fill=\"#fff\"\/><\/g>\n<g fill=\"none\" stroke=\"#5b0516\" stroke-width=\"5\" stroke-dasharray=\"0 10.5\"><circle cx=\"210\" cy=\"190\" r=\"86\"\/><circle cx=\"210\" cy=\"190\" r=\"80\"\/><\/g>\n<g stroke=\"#1c1718\" stroke-width=\"3\"><circle cx=\"222\" cy=\"182\" r=\"24\" fill=\"#e7e1e2\"\/><ellipse cx=\"176\" cy=\"216\" rx=\"14\" ry=\"8\" fill=\"#fff\" transform=\"rotate(-20 176 216)\"\/><ellipse cx=\"240\" cy=\"230\" rx=\"10\" ry=\"6\" fill=\"#fff\"\/><\/g>\n<path d=\"M380 140v160h210\" fill=\"none\" stroke=\"#6b6264\" stroke-width=\"2\"\/>\n<path d=\"M560 230v70\" fill=\"none\" stroke=\"#6b6264\" stroke-width=\"3\" stroke-dasharray=\"6 7\"\/>\n<path class=\"ibf-anim-sweep\" d=\"M380 300h90a90 90 0 0 0-6.1-32.6z\" fill=\"#8f0822\"\/>\n<path d=\"M380 300h162\" stroke=\"#1c1718\" stroke-width=\"6\"\/><path d=\"M558 300l-18-8v16z\" fill=\"#1c1718\" stroke=\"#1c1718\" stroke-width=\"3\"\/>\n<path d=\"M380 300l164-63.7\" stroke=\"#5b0516\" stroke-width=\"6\"\/><path d=\"M560 230l-19.6 0.2 6.4 15.3z\" fill=\"#5b0516\" stroke=\"#5b0516\" stroke-width=\"3\"\/>\n<circle cx=\"380\" cy=\"300\" r=\"6\" fill=\"#1c1718\"\/>\n<\/svg><\/div><figcaption class=\"ibf-f\">Phase angle reflects cell membrane health: low in malnutrition and cachexia, higher in healthy cells<\/figcaption><\/figure>\n<h3>Extracellular to total body water ratio (ECW\/TBW)<\/h3>\n<p>The ratio of extracellular water to total body water is a direct fluid-balance indicator. It shifts with oedema, inflammation and fluid overload \u2014 which makes it directly relevant to dialysis management, cardiac failure, and post-surgical monitoring. Assessing it properly requires multiple frequencies, because separating intracellular from extracellular water is exactly what low and high frequency current does differently.<\/p>\n<h3>Visceral fat area in cm\u00b2<\/h3>\n<p>A visceral fat <em>level<\/em> is adequate for wellness contexts. A visceral fat <em>area<\/em> in cm\u00b2 is a quantified value against a recognised clinical threshold \u2014 above 100 cm\u00b2 indicates elevated metabolic risk. For diabetology, endocrinology and metabolic clinics, the quantified area is the usable output.<\/p>\n<h2>Which Departments Use What<\/h2>\n<table>\n<thead>\n<tr>\n<th>Department<\/th>\n<th>Primary outputs used<\/th>\n<th>Typical application<\/th>\n<\/tr>\n<\/thead>\n<tbody>\n<tr>\n<td>Nephrology \/ dialysis<\/td>\n<td>ECW\/TBW, segmental water, Phase Angle<\/td>\n<td>Dry weight assessment, fluid overload monitoring<\/td>\n<\/tr>\n<tr>\n<td>Oncology<\/td>\n<td>Phase Angle, skeletal muscle mass<\/td>\n<td>Cachexia detection, nutritional status, treatment tolerance<\/td>\n<\/tr>\n<tr>\n<td>Diabetology \/ endocrinology<\/td>\n<td>Visceral fat area, SMM, body fat %<\/td>\n<td>Metabolic risk stratification, therapy response<\/td>\n<\/tr>\n<tr>\n<td>Critical care<\/td>\n<td>ECW\/TBW, Phase Angle<\/td>\n<td>Fluid status, nutritional assessment<\/td>\n<\/tr>\n<tr>\n<td>Geriatrics<\/td>\n<td>Segmental lean mass, SMM index<\/td>\n<td>Sarcopenia screening against AWGS criteria<\/td>\n<\/tr>\n<tr>\n<td>Bariatric \/ obesity clinic<\/td>\n<td>Body fat mass, visceral fat area, SMM<\/td>\n<td>Pre- and post-operative composition tracking<\/td>\n<\/tr>\n<tr>\n<td>Sports medicine \/ rehab<\/td>\n<td>Segmental lean analysis<\/td>\n<td>Post-injury asymmetry, rehabilitation progress<\/td>\n<\/tr>\n<tr>\n<td>Paediatrics<\/td>\n<td>Growth-referenced composition<\/td>\n<td>Nutritional assessment in children from age 3<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>Geriatrics deserves a specific note. Sarcopenia screening under the Asian Working Group for Sarcopenia criteria depends on appendicular skeletal muscle mass index, which requires reliable segmental limb measurement \u2014 not a whole-body figure. Our summary of the current criteria is at <a href=\"\/blog\/awgs-2025-sarcopenia-criteria-whats-changed-for-clinicians\">AWGS 2025 sarcopenia criteria for clinicians<\/a>.<\/p>\n<figure class=\"ibf ibf-bars\" data-ibf><figcaption class=\"ibf-t\">Frequencies decide which clinical outputs exist<\/figcaption><div class=\"ibf-bars\"><div class=\"ibf-br\"><span class=\"ibf-bl\">InBody 270S class (20 to 100 kHz)<\/span><span class=\"ibf-bv\">3\u00a0frequencies<\/span><span class=\"ibf-bt\"><span class=\"ibf-bf ibf-a-grow\" style=\"width:50%;--i:0\"><\/span><\/span><\/div><div class=\"ibf-br is-hi\"><span class=\"ibf-bl\">InBody 770S (1 to 1,000 kHz)<\/span><span class=\"ibf-bv\">6\u00a0frequencies<\/span><span class=\"ibf-bt\"><span class=\"ibf-bf ibf-a-grow\" style=\"width:100%;--i:1\"><\/span><\/span><\/div><\/div><div class=\"ibf-f\"><span>Six frequencies add phase angle, segmental ECW\/TBW and visceral fat area in cm\u00b2<\/span><\/div><\/figure>\n<h2>Why Frequency Count Is the Deciding Specification<\/h2>\n<p>For clinical work, the number of measurement frequencies is not a marketing tier. It determines which outputs are physically obtainable.<\/p>\n<ul>\n<li><strong>Three frequencies<\/strong> (20, 50, 100 kHz \u2014 InBody 270S class) gives segmental muscle and fat, visceral fat level, and basic water compartments. Suitable for outpatient screening, wellness and general clinical use.<\/li>\n<li><strong>Six frequencies<\/strong> (1 kHz to 1,000 kHz \u2014 InBody 770S) gives Phase Angle, segmental Phase Angle, segmental ECW\/TBW and visceral fat area in cm\u00b2. Required for nephrology, oncology, critical care and research.<\/li>\n<\/ul>\n<p>Low-frequency current travels primarily through extracellular fluid; higher frequencies penetrate cell membranes into intracellular fluid. Without a wide frequency range, the intracellular and extracellular compartments cannot be cleanly separated \u2014 and every fluid-status output depends on that separation.<\/p>\n<p>The InBody 970S extends this further with multi-frequency measurement up to 3 MHz and stores up to 100,000 measurements when patient IDs are entered, which matters for longitudinal departmental use.<\/p>\n<h2>Reproducibility: The Requirement Nobody Writes Into the Tender<\/h2>\n<p>In a clinical setting, the question is rarely &#8220;what is this patient&#8217;s body fat percentage today&#8221;. It is &#8220;has this changed since last week, and does that change reflect the intervention&#8221;.<\/p>\n<p>That makes reproducibility the governing requirement, and it depends on two things.<\/p>\n<p><strong>Fixed measurement start points.<\/strong> Tetrapolar 8-point tactile electrodes fix the measurement origin at the wrist and ankle every time. Without a fixed anatomical start point, small differences in hand or foot placement introduce variation that looks like clinical change.<\/p>\n<p><strong>No empirical estimation.<\/strong> If a device applies corrections based on age, sex or ethnicity, part of every result reflects population data rather than the patient. In a clinical context this is not a minor caveat \u2014 it means the device is partially reporting a demographic average back to you. InBody devices calculate directly from measured impedance per segment without these corrections.<\/p>\n<figure class=\"ibf ibf-illus\" data-ibf><div class=\"ibf-art ibf-a-up\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 640 400\" stroke-linecap=\"round\" stroke-linejoin=\"round\" role=\"img\" focusable=\"false\"><title>Tetrapolar 8-point electrodes fix the measurement start at wrist and ankle on every test<\/title>\n<circle cx=\"390\" cy=\"200\" r=\"165\" fill=\"#f6e4e9\"\/>\n<ellipse cx=\"380\" cy=\"356\" rx=\"110\" ry=\"9\" fill=\"#e7e1e2\"\/>\n<g transform=\"translate(380 55.2) scale(1.2)\"><path d=\"M-10 138L-12 226\" fill=\"none\" stroke=\"#1c1718\" stroke-width=\"22\"\/><path d=\"M-10 138L-12 226\" fill=\"none\" stroke=\"#fff\" stroke-width=\"16\"\/><path d=\"M10 138L12 226\" fill=\"none\" stroke=\"#1c1718\" stroke-width=\"22\"\/><path d=\"M10 138L12 226\" fill=\"none\" stroke=\"#fff\" stroke-width=\"16\"\/><path d=\"M-8 44H8Q24 44 24 60V126Q24 142 8 142H-8Q-24 142 -24 126V60Q-24 44 -8 44Z\" fill=\"#fff\" stroke=\"#1c1718\" stroke-width=\"3\"\/><path d=\"M-18 58L-47.4 138.8\" fill=\"none\" stroke=\"#1c1718\" stroke-width=\"22\"\/><path d=\"M-18 58L-47.4 138.8\" fill=\"none\" stroke=\"#fff\" stroke-width=\"16\"\/><path d=\"M18 58L47.4 138.8\" fill=\"none\" stroke=\"#1c1718\" stroke-width=\"22\"\/><path d=\"M18 58L47.4 138.8\" fill=\"none\" stroke=\"#fff\" stroke-width=\"16\"\/><circle cy=\"20\" r=\"20\" fill=\"#e7e1e2\" stroke=\"#1c1718\" stroke-width=\"3\"\/><path d=\"M-47.4 138.8L-18 58Q0 66 18 58L47.4 138.8M0 62V132L-12 228M0 132L12 228\" fill=\"none\" stroke=\"#8f0822\" stroke-width=\"3.5\" stroke-dasharray=\"7 7\" class=\"ibf-anim-flow\"\/><rect x=\"-46\" y=\"234\" width=\"92\" height=\"14\" rx=\"7\" fill=\"#fff\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"-47.4\" cy=\"129.8\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"-50.4\" cy=\"145.8\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"-5\" cy=\"241\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"-19\" cy=\"241\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"47.4\" cy=\"129.8\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"50.4\" cy=\"145.8\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"5\" cy=\"241\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><circle cx=\"19\" cy=\"241\" r=\"5.5\" fill=\"#8f0822\" stroke=\"#1c1718\" stroke-width=\"2.5\"\/><\/g>\n<g transform=\"translate(190 120)\"><circle r=\"34\" fill=\"#fff\" stroke=\"#1c1718\" stroke-width=\"3\"\/><path d=\"M-20 0C-14 -16 -6 -16 0 0S14 16 20 0\" fill=\"none\" stroke=\"#8f0822\" stroke-width=\"4\"\/><\/g><path d=\"M226 128Q262 140 313.1 211\" fill=\"none\" stroke=\"#6b6264\" stroke-width=\"2\" stroke-dasharray=\"2 6\"\/>\n\n<\/svg><\/div><figcaption class=\"ibf-f\">Tetrapolar 8-point electrodes fix the measurement start at wrist and ankle on every test<\/figcaption><\/figure>\n<h2>What to Put in the Specification Document<\/h2>\n<ol>\n<li>Direct Segmental Multi-frequency BIA with independent measurement of all five body segments.<\/li>\n<li>Tetrapolar 8-point tactile electrode system with fixed anatomical measurement start points.<\/li>\n<li>Minimum six measurement frequencies where Phase Angle or fluid analysis is required; minimum three otherwise.<\/li>\n<li>No empirical estimation based on age, sex or ethnicity \u2014 state this as a mandatory requirement, not a preference.<\/li>\n<li>Phase Angle output, whole-body and segmental, where clinically indicated.<\/li>\n<li>Segmental and whole-body ECW\/TBW ratio where fluid status is relevant.<\/li>\n<li>Visceral fat area reported in cm\u00b2, not level alone.<\/li>\n<li>Published validation against DEXA with study citation.<\/li>\n<li>Patient-ID-based data storage with export capability and result-sheet printing suited to medical records.<\/li>\n<li>Defined calibration schedule, service response time and authorised warranty support in India.<\/li>\n<li>Test duration compatible with OPD throughput \u2014 around 60 seconds for the measurement itself.<\/li>\n<li>Paediatric result sheet where the department tests children.<\/li>\n<\/ol>\n<h2>Installation and Practical Considerations<\/h2>\n<p>A few points that surface after purchase rather than before:<\/p>\n<ul>\n<li><strong>Siting.<\/strong> The device needs a level, stable floor and should not sit adjacent to sources of significant electrical interference.<\/li>\n<li><strong>Protocol standardisation.<\/strong> Departments that get clean longitudinal data write a fixed testing protocol \u2014 time of day, position relative to dialysis or meals, hydration guidance \u2014 and enforce it. Departments that do not, generate noise.<\/li>\n<li><strong>Staff training.<\/strong> The outputs are only as useful as the clinician&#8217;s familiarity with them. Phase Angle in particular is underused simply because staff have not been briefed on interpretation.<\/li>\n<li><strong>Contraindications.<\/strong> As with any device passing a current through the body, follow the manufacturer&#8217;s guidance regarding patients with implanted electronic medical devices such as pacemakers.<\/li>\n<\/ul>\n<p>InBody has 3,000+ installations across India, at hospitals, clinics and gyms including AIIMS Delhi, PGI Chandigarh and Tata Memorial Mumbai, which makes peer reference checks straightforward during procurement.<\/p>\n<figure class=\"ibf ibf-stat\" data-ibf><span class=\"ibf-sc ibf-a-up\"><svg class=\"ibf-i\" viewBox=\"0 0 24 24\" aria-hidden=\"true\" focusable=\"false\"><path d=\"M6.8 2l2 1 .8 2.1L11 6.8l2 1.1 2.8.7.7.6 2.1-.3 3.4-.7-2.8 2.8-.6 1-1.4-2-.7 2.3-1.5.4-1.3 1.3-2 1.8-1.2 1-.2 1.8-.2 1.8-1.5 1.5-.9-.9-.9-2.1-1-2.4-.5-2.1-.1-1.4H3.7L2.5 12l-.2-.8 1.4-.6.4-1 1.7-2.1.7-1.7-.4-1.7z\"\/><\/svg><\/span><span class=\"ibf-stv ibf-a-up\">3,000+<\/span><figcaption class=\"ibf-a-up\"><span class=\"ibf-stl\">InBody installations across India, including AIIMS Delhi, PGI Chandigarh and Tata Memorial Mumbai<\/span><\/figcaption><\/figure>\n<p><script type=\"application\/ld+json\">\n{\n  \"@context\": \"https:\/\/schema.org\",\n  \"@type\": \"FAQPage\",\n  \"mainEntity\": [\n    {\n      \"@type\": \"Question\",\n      \"name\": \"What specification should a hospital require in a body composition analyzer?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Direct Segmental Multi-frequency BIA measuring all five body segments, tetrapolar 8-point tactile electrodes with fixed anatomical start points, and no empirical estimation based on age, sex or ethnicity. Six measurement frequencies are required where Phase Angle or fluid analysis is needed. Visceral fat should be reported as an area in cm\u00b2 rather than a level alone.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"Why does a hospital need six frequencies rather than three?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Frequency range determines which outputs are physically obtainable. Low-frequency current travels through extracellular fluid while higher frequencies penetrate cell membranes into intracellular fluid. Without a wide range, the two compartments cannot be cleanly separated, and every fluid-status output \u2014 including Phase Angle and ECW\/TBW ratio \u2014 depends on that separation.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"What is Phase Angle used for clinically?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Phase Angle reflects cell membrane integrity and cellular health, derived from the ratio of reactance to resistance. Lower values indicate cellular stress seen in malnutrition, cancer cachexia and chronic disease. It is used in oncology, nephrology and critical care to assess nutritional status and as a prognostic indicator alongside other clinical findings.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"Can a body composition analyzer be used for sarcopenia screening?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Yes, provided it measures limb segments independently. Sarcopenia screening under Asian Working Group for Sarcopenia criteria depends on appendicular skeletal muscle mass index, which requires reliable segmental measurement of the arms and legs rather than a whole-body muscle figure. Whole-body devices cannot produce the required index.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"How long does a clinical body composition test take?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"The measurement itself takes roughly 60 seconds with the patient standing and holding the hand electrodes, fully clothed and with no injection or radiation involved. Including patient registration and result review, most outpatient departments complete the encounter within about 10 minutes, which is compatible with normal OPD throughput.\"\n      }\n    }\n  ]\n}\n<\/script><\/p>\n<h2>Specify It Around Your Departments<\/h2>\n<p>The right analyser for a dialysis unit is not the right analyser for a wellness OPD. The specification should follow the clinical question each department needs answered.<\/p>\n<p><strong>Request a specification-matched quote for your hospital or clinic at <a href=\"https:\/\/www.inbody.in\/contact-us.php?ref=blog&#038;cta=inline-4599\">inbody.in\/contact-us.php<\/a><\/strong>, or contact the clinical team through <a href=\"https:\/\/www.inbody.in\/contact-us.php?ref=blog&#038;cta=inline-4599\">our enquiry form<\/a> to discuss departmental requirements. Related reading: <a href=\"\/blog\/phase-angle-explained-what-is-a-good-phase-angle-and-why-it-matters\">phase angle explained<\/a>, <a href=\"\/blog\/awgs-2025-sarcopenia-criteria-whats-changed-for-clinicians\">AWGS sarcopenia criteria<\/a>, and <a href=\"https:\/\/www.inbody.in\/blog\/bca-machine-complete-guide-india\">how to choose a body composition analyser<\/a>.<\/p>\n<h2 id=\"sources\">Sources<\/h2>\n<ol>\n<li><a href=\"https:\/\/pubmed.ncbi.nlm.nih.gov\/22698802\/\" target=\"_blank\" rel=\"noopener\">Norman K, Stob\u00e4us N, Pirlich M, et al. Bioelectrical phase angle and impedance vector analysis&#8211;clinical relevance and applicability of impedance parameters. Clin Nutr. 2012;31(6):854-61.<\/a><\/li>\n<li><a href=\"https:\/\/pubmed.ncbi.nlm.nih.gov\/33581951\/\" target=\"_blank\" rel=\"noopener\">Arab A, Karimi E, Vingrys K, et al. Is phase angle a valuable prognostic tool in cancer patients&#8217; survival? A systematic review and meta-analysis of available literature. Clin Nutr. 2021;40(5):3182-3190.<\/a><\/li>\n<li><a href=\"https:\/\/pubmed.ncbi.nlm.nih.gov\/32033882\/\" target=\"_blank\" rel=\"noopener\">Chen LK, Woo J, Assantachai P, et al. Asian Working Group for Sarcopenia: 2019 Consensus Update on Sarcopenia Diagnosis and Treatment. J Am Med Dir Assoc. 2020;21(3):300-307.e2.<\/a><\/li>\n<li><a href=\"https:\/\/pubmed.ncbi.nlm.nih.gov\/12419927\/\" target=\"_blank\" rel=\"noopener\">Examination Committee of Criteria for &#8216;Obesity Disease&#8217; in Japan; Japan Society for the Study of Obesity. New criteria for &#8216;obesity disease&#8217; in Japan. Circ J. 2002;66(11):987-92.<\/a><\/li>\n<\/ol>\n","protected":false},"excerpt":{"rendered":"<p>Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.<\/p>\n","protected":false},"author":4,"featured_media":5440,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_uf_show_specific_survey":0,"_uf_disable_surveys":false,"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_hide_featured_image":false,"_jetpack_featured_image_duplicate_dismissed":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[9,13,38],"tags":[814,812,817,815,816,813,818,68],"class_list":["post-4599","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-body","category-health","category-medical","tag-ecw-tbw","tag-hospital-body-composition-analyzer","tag-inbody-770s","tag-nephrology","tag-oncology","tag-phase-angle","tag-procurement","tag-sarcopenia"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.9 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Body Composition Analyzer for Hospitals: Clinical Spec Guide<\/title>\n<meta name=\"description\" content=\"Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Body Composition Analyzer for Hospitals: Clinical Spec Guide\" \/>\n<meta property=\"og:description\" content=\"Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india\" \/>\n<meta property=\"og:site_name\" content=\"InBody India Blog\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/inbodyindia\/\" \/>\n<meta property=\"article:published_time\" content=\"2026-08-17T12:57:27+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-11-17T03:41:21+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/inbody-770-clinical-analyzer.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"435\" \/>\n\t<meta property=\"og:image:height\" content=\"900\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Udbhav Kamath\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Udbhav Kamath\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"7 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india\"},\"author\":{\"name\":\"Udbhav Kamath\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/#\\\/schema\\\/person\\\/a909d8aa1b4c3eb03ea10dd9398cde45\"},\"headline\":\"Body Composition Analyzer for Hospitals: Clinical Spec Guide\",\"datePublished\":\"2026-08-17T12:57:27+00:00\",\"dateModified\":\"2026-11-17T03:41:21+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india\"},\"wordCount\":1373,\"commentCount\":1,\"publisher\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/#organization\"},\"image\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/wp-content\\\/uploads\\\/2026\\\/08\\\/4599.webp\",\"keywords\":[\"ECW\\\/TBW\",\"hospital body composition analyzer\",\"InBody 770S\",\"nephrology\",\"oncology\",\"Phase Angle\",\"procurement\",\"sarcopenia\"],\"articleSection\":[\"Body Composition\",\"Health\",\"Medical\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india\",\"url\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india\",\"name\":\"Body Composition Analyzer for Hospitals: Clinical Spec Guide\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/wp-content\\\/uploads\\\/2026\\\/08\\\/4599.webp\",\"datePublished\":\"2026-08-17T12:57:27+00:00\",\"dateModified\":\"2026-11-17T03:41:21+00:00\",\"description\":\"Phase Angle, ECW\\\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#primaryimage\",\"url\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/wp-content\\\/uploads\\\/2026\\\/08\\\/4599.webp\",\"contentUrl\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/wp-content\\\/uploads\\\/2026\\\/08\\\/4599.webp\",\"width\":1600,\"height\":900,\"caption\":\"Illustration for: Body Composition Analyzer for Hospitals: Clinical Spec Guide\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/body-composition-analyzer-for-hospitals-india#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Body Composition Analyzer for Hospitals: Clinical Spec Guide\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/#website\",\"url\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/\",\"name\":\"InBody India Blog\",\"description\":\"See What You&#039;re Made Of.\",\"publisher\":{\"@id\":\"https:\\\/\\\/www.inbody.in\\\/#organization\"},\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Organization\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/#organization\",\"name\":\"InBody India\",\"legalName\":\"InBody India Private Limited\",\"url\":\"https:\\\/\\\/www.inbody.in\\\/\",\"logo\":{\"@type\":\"ImageObject\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/#logo\",\"url\":\"https:\\\/\\\/www.inbody.in\\\/images\\\/InBody-logo-with-tagline.png\",\"width\":400,\"height\":153},\"image\":\"https:\\\/\\\/www.inbody.in\\\/images\\\/InBody-logo-with-tagline.png\",\"telephone\":\"+91-80975-11852\",\"email\":\"inquiry_india@inbody.com\",\"address\":{\"@type\":\"PostalAddress\",\"streetAddress\":\"SF-B-03, B Wing, 2nd Floor, Art Guild House, Phoenix Market City, L.B.S Marg, Kurla West\",\"addressLocality\":\"Mumbai\",\"addressRegion\":\"Maharashtra\",\"postalCode\":\"400070\",\"addressCountry\":\"IN\"},\"contactPoint\":[{\"@type\":\"ContactPoint\",\"telephone\":\"+91-80975-11852\",\"contactType\":\"sales\",\"areaServed\":\"IN\",\"availableLanguage\":[\"English\",\"Hindi\"]},{\"@type\":\"ContactPoint\",\"telephone\":\"+91-86574-19175\",\"contactType\":\"customer support\",\"areaServed\":\"IN\",\"availableLanguage\":[\"English\",\"Hindi\"]}],\"sameAs\":[\"https:\\\/\\\/www.facebook.com\\\/inbodyindia\",\"https:\\\/\\\/www.instagram.com\\\/inbodyindia\\\/\",\"https:\\\/\\\/www.linkedin.com\\\/company\\\/inbody-india\",\"https:\\\/\\\/www.youtube.com\\\/@inbodyindiaofficial\"],\"parentOrganization\":{\"@type\":\"Organization\",\"name\":\"InBody Co., Ltd.\",\"url\":\"https:\\\/\\\/inbody.com\\\/\"}},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/www.inbody.in\\\/blog\\\/#\\\/schema\\\/person\\\/a909d8aa1b4c3eb03ea10dd9398cde45\",\"name\":\"Udbhav Kamath\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g\",\"caption\":\"Udbhav Kamath\"},\"description\":\"Udbhav Kamath manages content and digital at InBody India, the official Indian subsidiary of InBody Co., Ltd.\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Body Composition Analyzer for Hospitals: Clinical Spec Guide","description":"Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india","og_locale":"en_US","og_type":"article","og_title":"Body Composition Analyzer for Hospitals: Clinical Spec Guide","og_description":"Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.","og_url":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india","og_site_name":"InBody India Blog","article_publisher":"https:\/\/www.facebook.com\/inbodyindia\/","article_published_time":"2026-08-17T12:57:27+00:00","article_modified_time":"2026-11-17T03:41:21+00:00","og_image":[{"width":435,"height":900,"url":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/inbody-770-clinical-analyzer.jpg","type":"image\/jpeg"}],"author":"Udbhav Kamath","twitter_card":"summary_large_image","twitter_misc":{"Written by":"Udbhav Kamath","Est. reading time":"7 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#article","isPartOf":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india"},"author":{"name":"Udbhav Kamath","@id":"https:\/\/www.inbody.in\/blog\/#\/schema\/person\/a909d8aa1b4c3eb03ea10dd9398cde45"},"headline":"Body Composition Analyzer for Hospitals: Clinical Spec Guide","datePublished":"2026-08-17T12:57:27+00:00","dateModified":"2026-11-17T03:41:21+00:00","mainEntityOfPage":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india"},"wordCount":1373,"commentCount":1,"publisher":{"@id":"https:\/\/www.inbody.in\/#organization"},"image":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#primaryimage"},"thumbnailUrl":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/4599.webp","keywords":["ECW\/TBW","hospital body composition analyzer","InBody 770S","nephrology","oncology","Phase Angle","procurement","sarcopenia"],"articleSection":["Body Composition","Health","Medical"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india","url":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india","name":"Body Composition Analyzer for Hospitals: Clinical Spec Guide","isPartOf":{"@id":"https:\/\/www.inbody.in\/blog\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#primaryimage"},"image":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#primaryimage"},"thumbnailUrl":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/4599.webp","datePublished":"2026-08-17T12:57:27+00:00","dateModified":"2026-11-17T03:41:21+00:00","description":"Phase Angle, ECW\/TBW and visceral fat area separate clinical from fitness devices. Department-wise outputs plus a 12-point tender specification.","breadcrumb":{"@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#primaryimage","url":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/4599.webp","contentUrl":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/4599.webp","width":1600,"height":900,"caption":"Illustration for: Body Composition Analyzer for Hospitals: Clinical Spec Guide"},{"@type":"BreadcrumbList","@id":"https:\/\/www.inbody.in\/blog\/body-composition-analyzer-for-hospitals-india#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.inbody.in\/blog\/"},{"@type":"ListItem","position":2,"name":"Body Composition Analyzer for Hospitals: Clinical Spec Guide"}]},{"@type":"WebSite","@id":"https:\/\/www.inbody.in\/blog\/#website","url":"https:\/\/www.inbody.in\/blog\/","name":"InBody India Blog","description":"See What You&#039;re Made Of.","publisher":{"@id":"https:\/\/www.inbody.in\/#organization"},"potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.inbody.in\/blog\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Organization","@id":"https:\/\/www.inbody.in\/#organization","name":"InBody India","legalName":"InBody India Private Limited","url":"https:\/\/www.inbody.in\/","logo":{"@type":"ImageObject","@id":"https:\/\/www.inbody.in\/#logo","url":"https:\/\/www.inbody.in\/images\/InBody-logo-with-tagline.png","width":400,"height":153},"image":"https:\/\/www.inbody.in\/images\/InBody-logo-with-tagline.png","telephone":"+91-80975-11852","email":"inquiry_india@inbody.com","address":{"@type":"PostalAddress","streetAddress":"SF-B-03, B Wing, 2nd Floor, Art Guild House, Phoenix Market City, L.B.S Marg, Kurla West","addressLocality":"Mumbai","addressRegion":"Maharashtra","postalCode":"400070","addressCountry":"IN"},"contactPoint":[{"@type":"ContactPoint","telephone":"+91-80975-11852","contactType":"sales","areaServed":"IN","availableLanguage":["English","Hindi"]},{"@type":"ContactPoint","telephone":"+91-86574-19175","contactType":"customer support","areaServed":"IN","availableLanguage":["English","Hindi"]}],"sameAs":["https:\/\/www.facebook.com\/inbodyindia","https:\/\/www.instagram.com\/inbodyindia\/","https:\/\/www.linkedin.com\/company\/inbody-india","https:\/\/www.youtube.com\/@inbodyindiaofficial"],"parentOrganization":{"@type":"Organization","name":"InBody Co., Ltd.","url":"https:\/\/inbody.com\/"}},{"@type":"Person","@id":"https:\/\/www.inbody.in\/blog\/#\/schema\/person\/a909d8aa1b4c3eb03ea10dd9398cde45","name":"Udbhav Kamath","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/b30b5fc1dd126c5aa892141bd02ab09b528e759b43e89658cd0f5284981e8294?s=96&d=mm&r=g","caption":"Udbhav Kamath"},"description":"Udbhav Kamath manages content and digital at InBody India, the official Indian subsidiary of InBody Co., Ltd."}]}},"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/www.inbody.in\/blog\/wp-content\/uploads\/2026\/08\/4599.webp","_links":{"self":[{"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/posts\/4599","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/comments?post=4599"}],"version-history":[{"count":2,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/posts\/4599\/revisions"}],"predecessor-version":[{"id":5264,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/posts\/4599\/revisions\/5264"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/media\/5440"}],"wp:attachment":[{"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/media?parent=4599"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/categories?post=4599"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.inbody.in\/blog\/wp-json\/wp\/v2\/tags?post=4599"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}